Mixed signal controller for an experimental magnetotherapy device

Authors

  • Paweł Pawłowski Division of Signal Processing and Electronic Systems, Institute of Automatic Control and Robotics, Poznan University of Technology, Poland https://orcid.org/0000-0001-5373-5148
  • Marek Portalski

DOI:

https://doi.org/10.24425/ijet.2025.153557

Abstract

The article presents the design of a controller for an experimental magnetotherapy device. The controller realizes precise, digital generation of a 3-phase sinusoidal signal with a frequency of 1 Hz to 5 kHz and slowly variable frequency modulation of 1 Hz to 5 Hz, which is a novelty in the field of magnetotherapy. The structure and operation of the analog filter path and the regulation of the amplitude and the DC component are discussed. A fast STM32F334 microcontroller was used to generate the signal, and an ATmega2560 microcontroller was used to operate the user interface with a color touch display. Laboratory and therapeutic tests confirmed proper operation of the controller.

Published

2025-04-07

How to Cite

Pawłowski, Paweł, and Marek Portalski. “Mixed Signal Controller for an Experimental Magnetotherapy Device”. International Journal of Electronics and Telecommunications, vol. 71, no. 1, Apr. 2025, doi:10.24425/ijet.2025.153557.

Issue

Section

Artykuły