Effect of the illuminance meter used on the results of color light pollution measurements

Authors

  • Przemysław Tabaka Technical University of Lodz, at the Institute of Electrical Engineering, Poland https://orcid.org/0000-0001-6716-5043
  • Justyna Wtorkiewicz Regional Office of Measures in Lodz, at theLaboratory of Physicochemistry, Poland

DOI:

https://doi.org/10.24425/ijet.2024.152089

Abstract

The paper presents the results of measurements of illuminance. Twenty-two types of illuminance meter with known spectral sensitivity curves of photometric heads and eleven lamps emitting colored light were used. Laboratory measurements were conducted at a characteristic point, with a focus on assessing intrusive light, with a value of 3 lx. The considerations were supplemented with simulated illuminance calculations. For most illuminance meters, correlations were observed between the measurement results and the calculations. This allows the conclusion that, especially in the case of colored light, one of the main factors affecting the measurement result is the spectral sensitivity curve of the photometric head. The greatest discrepancies in the obtained illuminance values were found in the case of lamps emitting blue light. The paper highlights the fact that measuring low illuminance values produced by lamps emitting colored light is not an easy task. Depending on the illuminance meter used, conclusions regarding the assessment of intrusive light may vary. The aim of the paper is to indicate that not all calibrated illuminance meters are suitable for this assessment.

Published

2024-11-25

How to Cite

Tabaka, Przemysław, and Justyna Wtorkiewicz. “Effect of the Illuminance Meter Used on the Results of Color Light Pollution Measurements”. International Journal of Electronics and Telecommunications, vol. 70, no. 4, Nov. 2024, doi:10.24425/ijet.2024.152089.

Issue

Section

Artykuły