Design and validation of a forehead pulse oximeter for military medical evacuation
DOI:
https://doi.org/10.24425/mms.2025.154676Abstract
This article presents the design, implementation, and validation of an SpO2 sensor prototype, developed as part of a military medical evacuation system. The reflectance method was chosen for the measurement of SpO2, allowing readings to be taken from any location on the body, unlike the traditional transmissive method. The sensor is designed as a headband worn on the forehead, a well-perfused area, ensuring that the placement does not hinder the mobility of the soldier. The sensor construction utilized an SFH7072 optical module, an AFE4410 integrated circuit, and an STM32L5 microcontroller. To obtain accurate SpO2 readings, signal processing techniques, including filtering, were applied, along with the development of algorithms to calculate SpO2 from photoplethysmographic (PPG) signals. The prototype underwent validation tests using a comprehensive experimental setup. The prototype achieved an ARMS (root mean square difference) value of 3.0%, meeting the ISO 80601-2-61 standard for reflectance sensors, which recommends an 𝐴RMS ≤ 3.5%.References
[1] Lubkowski, P., Krygier, J., Sondej, T., Dobrowolski, A. P., Apiecionek, L., Znaniecki, W., Oskwarek, P. (2023). Decision Support System Proposal for Medical Evacuations in Military Operations. Sensors (Basel, Switzerland), 23(11), 5144. https://doi.org/10.3390/s23115144
[2] Krygier, J., Lubkowski, P., Maslanka, K., Dobrowolski, A. P., Mrozek, T., Znaniecki, W. & Oskwarek, P. (2024). Smart Medical Evacuation Support System for the Military. Sensors (Basel, Switzerland), 24(14), 4581. https://doi.org/10.3390/s24144581
[3] Bard, D. M., Joseph, J. I. & van Helmond, N. (2019). Cuff-Less Methods for Blood Pressure Telemonitoring. Frontiers in Cardiovascular Medicine, 6, 40. https://doi.org/10.3389/fcvm.2019.00040
[4] Yao, L. P. & Pan, Z. L. (2021). Cuff-less blood pressure estimation from photoplethysmography signal and electrocardiogram. Physical and Engineering Sciences in Medicine, 44(2), 397-408. https://doi.org/10.1007/s13246-021-00989-1
[5] Chu, M., Nguyen, T., Pandey, V., Zhou, Y., Pham, H. N., Bar-Yoseph, R., Radom-Aizik, S., Jain, R., Cooper, D. M. & Khine, M. (2019). Respiration rate and volume measurements using wearable strain sensors. NPJ Digital Medicine, 2, 8. https://doi.org/10.1038/s41746-019-0083-3
[6] Hussain, T., Ullah, S., Fernández-García, R. & Gil, I. (2023). Wearable Sensors for Respiration Monitoring: A Review. Sensors, 23(17), 7518. https://doi.org/10.3390/s23177518
[7] Szmajda, M., Chylinski, M., Sacha, J. & Mroczka, J. (2023). Three methods for determining the respiratory waves from ECG (Part I). Metrology and Measurement Systems, 30 (4), 821-837. https://doi.org/10.24425/mms.2023.147956
[8] Szmajda, M., Chylinski, M., Sacha, J. & Mroczka, J. (2024). Three methods for determining the respiratory waves from ECG (Part II). Metrology and Measurement Systems, 31 (1), 51-71. https://doi.org/10.24425/mms.2024.148544
[9] Chan, E. D., Chan, M. M. & Chan, M. M. (2013). Pulse oximetry: understanding its basic principles facilitates appreciation of its limitations. Respiratory Medicine, 107(6), 789-799. https://doi.org/10.1016/j.rmed.2013.02.004
[10] Nitzan, M., Romem, A. & Koppel, R. (2014). Pulse oximetry: fundamentals and technology update. Medical Devices: Evidence and Research (Auckland, N.Z.), 7, 231-239. https://doi.org/10.2147/MDER.S47319
[11] Tekin, K., Karadogan, M., Gunaydin, S. & Kismet, K. (2023). Everything about Pulse Oximetry-Part 1: History, Principles, Advantages, Limitations, Inaccuracies, Cost Analysis, the Level of Knowledge About Pulse Oximeter Among Clinicians, and Pulse Oximetry Versus Tissue Oximetry. Journal of Intensive Care Medicine, 38(9), 775-784. https://doi.org/10.1177/08850666231185752
[12] Tekin, K., Karadogan, M., Gunaydin, S. & Kismet, K. (2023). Everything About Pulse Oximetry - Part 2: Clinical Applications, Portable/Wearable Pulse Oximeters, Remote Patient Monitoring, and Recent Advances. Journal of Intensive Care Medicine, 38(10), 887-896. https://doi.org/10.1177/08850666231189175
[13] Indrawan, J., Solihat, Y., Wijaya, D.W., Nasution, A. H. & Sembiring, S. (2020). Pulse Oximetry Effectivity for Measuring Oxygen Saturation in Hypovolemic Shock Patient Compared to Blood Gas Analysis at Rsup H. Adam Malik Medan. International Journal of Scientific and Research Publications, 10(8). http://dx.doi.org/10.29322/IJSRP.10.08.2020.p10499
[14] Tamura T. (2019). Current progress of photoplethysmography and SpO2 for health monitoring. Biomedical Engineering Letters, 9(1), 21-36. https://doi.org/10.1007/s13534-019-00097-w
[15] Allen J. (2007). Photoplethysmography and its application in clinical physiological measurement. Physiological Measurement, 28(3), R1-R39. https://doi.org/10.1088/0967-3334/28/3/R01
[16] Motta, L. P., Silva, P. P. F. D., Borguezan, B. M., Amaral, J. L. M. D., Milagres, L. G., Bóia, M. N., Ferraz, M. R., Mogami, R., Nunes, R. A. & Melo, P. L. (2021). An emergency system for monitoring pulse oximetry, peak expiratory flow, and body temperature of patients with COVID-19 at home: Development and preliminary application. PloS One, 16(3), e0247635. https://doi.org/10.1371/journal.pone.0247635
[17] Mackway-Jones, K., Marsden, J. & Windle, J. (Eds.) (2013). Emergency Triage: Manchester Triage Group. John Wiley & Sons. https://doi.org/10.1002/9781118299029
[18] Gilboy, N., Tanabe, P., Travers, D. & Rosenau, A. M. (2011). Emergency Severity Index (ESI): A Triage Tool for Emergency Department Care, Version 4. Implementation Handbook 2012 Edition. AHRQ Publication.
[19] Royal College of Physicians (2017). National Early Warning Score (NEWS) 2: Standardising the Assessment of Acute-illness Severity in the NHS. Updated report of a working party. London.
[20] Smith, G. B., Prytherch, D. R., Meredith, P., Schmidt, P. E. & Featherstone, P. I. (2013). The ability of the National Early Warning Score (NEWS) to discriminate patients at risk of early cardiac arrest, unanticipated intensive care unit admission, and death. Resuscitation, 84(4), 465-470. https://doi.org/10.1016/j.resuscitation.2012.12.016
[21] Jubran A. (2015). Pulse oximetry. Critical Care, 19(1), 272. https://doi.org/10.1186/s13054-015-0984-8
[22] Butler, F. K., Bennett, B. & Wedmore, C. I. (2017). Tactical Combat Casualty Care and Wilderness Medicine: Advancing Trauma Care in Austere Environments. Emergency Medicine Clinics of North America, 35(2), 391-407. https://doi.org/10.1016/j.emc.2016.12.005
[23] Azudin, K., Gan, K. B., Jaafar, R. & Ja’afar, M. H. (2023). The Principles of Hearable Photoplethysmography Analysis and Applications in Physiological Monitoring - A Review. Sensors, 23(14), 6484. https://doi.org/10.3390/s23146484
[24] Kumar, Ch., Mullagiri, M., Maruthy, K., Siva Kumar, A.V. & Kuppusamy, M. (2021). Association of heart rate variability measured by RR interval from ECG and pulse to pulse interval from photoplethysmography, Clinical Epidemiology and Global Health, 100698. https://doi.org/10.1016/j.cegh.2021.100698
[25] Polak, A. G., Klich, B., Saganowski, S., Prucnal, M. A. & Kazienko, P. (2022). Processing Photoplethysmograms Recorded by Smartwatches to Improve the Quality of Derived Pulse Rate Variability. Sensors, 22(18), 7047. https://doi.org/10.3390/s22187047
[26] Prucnal, M. A, Polak, A. G. & Kazienko, P. (2025), Improving the quality of pulse rate variability derived from wearable devices using adaptive, spectrum and nonlinear filtering, Biomedical Signal Processing and Control, vol. 102, 107336. https://doi.org/10.1016/j.bspc.2024.107336
[27] Schoffl, J., Pozzato, I., Rodrigues, D., Arora, M. & Craig, A. (2023). Pulse rate variability: An alternative to heart rate variability in adults with spinal cord injury. Psychophysiology, 60(11), e14356. https://doi.org/10.1111/psyp.14356
[28] Antali, F., Kulin, D., Lucz, K. I., Szabó, B., Szűcs, L., Kulin, S. & Miklós, Z. (2021). Multimodal Assessment of the Pulse Rate Variability Analysis Module of a Photoplethysmography-Based Telemedicine System. Sensors, 21(16), 5544. https://doi.org/10.3390/s21165544
[29] Bachir, W. (2023). Diffuse transmittance visible spectroscopy using smartphone flashlight for photoplethysmography and vital signs measurements. Spectrochimica Acta. Part A, Molecular and Biomolecular Spectroscopy, 303, 123181. https://doi.org/10.1016/j.saa.2023.123181
[30] Allen J. & Kyriacou P. (2021). Photoplethysmography: Technology, Signal Analysis and Applications, Academic Press. https://doi.org/10.1016/C2020-0-00098-8
[31] Kozioł, M., Piech. P., Maciejewski, M. & Surtel, W. (2019). The latest applications of photoplethysmography, Acta Angiologica, 25(1), 28-34. https://doi.org/10.5603/AA.2019.0005
[32] Santos, M., Vollam, S., Pimentel, M. A., Areia, C., Young, L., Roman, C., Ede, J., Piper, P., King, E., Harford, M., Shah, A., Gustafson, O., Tarassenko, L. & Watkinson, P. (2022). The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study. Journal of Medical Internet Research, 24(2), e28890. https://doi.org/10.2196/28890
[33] Kałamajska, E., Misiurewicz, J. & Weremczuk, J. (2022). Wearable Pulse Oximeter for Swimming Pool Safety. Sensors, 22(10), 3823. https://doi.org/10.3390/s22103823
[34] Azhari, A., Yoshimoto, S., Nezu, T., Iida, H., Ota, H., Noda, Y., ... & Morii, K. (2017, October). A patch-type wireless forehead pulse oximeter for SpO 2 measurement. In 2017 IEEE Biomedical Circuits and Systems Conference (BioCAS) (pp. 1-4). IEEE. https://doi.org/10.1109/BIOCAS.2017.8325557
[35] Hummler, H. D., Engelmann, A., Pohlandt, F., Högel, J. & Franz, A. R. (2006). Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value? Intensive Care Medicine, 32, 1428-1431. https://doi.org/10.1007/s00134-006-0254-y
[36] Seifi, S., Khatony, A., Moradi, G., Abdi, A. & Najafi, F. (2018). Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: comparison of finger, toe, forehead and earlobe probes. BMC Nursing, 17, 1-7. https://doi.org/10.1186/s12912-018-0283-1
[37] Longmore, S. K., Lui, G. Y., Naik, G., Breen, P. P., Jalaludin, B. & Gargiulo, G. D. (2019). A comparison of reflective photoplethysmography for detection of heart rate, blood oxygen saturation, and respiration rate at various anatomical locations. Sensors, 19(8), 1874. https://doi.org/10.3390/s19081874
[38] Přibil, J., Přibilová, A. & Frollo, I. (2020). Comparative measurement of the PPG signal on different human body positions by sensors working in reflection and transmission modes. Engineering Proceedings, 2(1), 69. https://doi.org/10.3390/ecsa-7-08204
[39] Kelly, K. L., Carlson, A. R., Allison, T. G. & Johnson, B. D. (2020). A comparison of finger and forehead pulse oximeters in heart failure patients during maximal exercise. Heart & Lung, 49(3), 259-264. https://doi.org/10.1016/j.hrtlng.2019.10.012
[40] Sondej, T. & Zawadzka, S. (2021). Influence of cuff pressures of automatic sphygmomanometers on pulse oximetry measurements. Measurement, 187, 110329. https://doi.org/10.1016/j.measurement.2021.110329
[41] Masimo (n.d.). LNCS®TFA-1™Disposable Forehead Sensor. Retrieved on June 18, 2024, from https://techdocs.masimo.com/products/sensor/lncs-tfa-1-disposable-forehead-sensor/
[42] Medtronic (n.d.). Nellcor™Max-Fast™forehead SpO2 sensor. Retrieved on June 18, 2024, from https://www.medtronic.com/covidien/en-us/products/pulse-oximetry/nellcor-spo2-forehead-sensor.html
[43] Stockwell, S. J., Kwok, T. C., Morgan, S. P., Sharkey, D. & Hayes-Gill, B. R. (2023). Forehead monitoring of heart rate in neonatal intensive care. Frontiers in Physiology, 14, 1127419. https://doi.org/10.3389/fphys.2023.1127419
[44] Liu, S. H., Liu, H. C., Chen, W. & Tan, T. H. (2020). Evaluating quality of photoplethysmographic signal on wearable forehead pulse oximeter with supervised classification approaches. IEEE Access, 8, 185121-185135. https://doi.org/10.1109/ACCESS.2020.3029842
[45] Lee, J., Matsumura, K., Yamakoshi, K. I., Rolfe, P., Tanaka, S. & Yamakoshi, T. (2013, July). Comparison between red, green and blue light reflection photoplethysmography for heart rate monitoring during motion. In 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (pp. 1724-1727), IEEE. https://doi.org/10.1109/embc.2013.6609852
[46] Kirson, L. E. & Koltes-Edwards, R. (2010). Pulse Oximetry. Anesthesia Secrets E-Book, 168.
[47] Elgendi, M., Norton, I., Brearley, M., Abbott, D. & Schuurmans, D. (2013). Systolic peak detection in acceleration photoplethysmograms measured from emergency responders in tropical conditions. PLOS One, 8(10), e76585. https://doi.org/10.1371/journal.pone.0076585
[48] U.S. Food and Drug Administration (2013, March). Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/pulse-oximeters-premarket-notification-submissions-510ks-guidance-industry-and-food-and-drug
[49] Ludbrook, J. (2010). Confidence in Altman-Bland plots: A critical review of the method of differences. Clinical and Experimental Pharmacology and Physiology, 37(2), 143-149. https://doi.org/10.1111/j.1440-1681.2009.05288.x
[50] Louie, A., Feiner, J. R., Bickler, P. E., Rhodes, L., Bernstein, M. & Lucero, J. (2018). Four types of pulse oximeters accurately detect hypoxia during low perfusion and motion. Anesthesiology, 128(3), 520-530. https://doi.org/10.1097/aln.0000000000002002